Z-0887-2017 Class II Terminated

Recalled by Amendia, Inc — Marietta, GA

FDA device recall Z-0887-2017 was initiated by Amendia, Inc on Nov 22, 2016 and is designated Class II. Reason for recall: There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw. The recall status is terminated (terminated Jul 31, 2017). Affected quantity: 96 units.

Recall details

Product Type
Devices
Report Date
Dec 28, 2016
Initiation Date
Nov 22, 2016
Termination Date
Jul 31, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
96 units

Product description

Black Widow 6.6mm x 20mm Screw, Part Number TBS020; Black Widow 6.6mm x 25mm Screw, Part Number TBS025; Black Widow 6.6mm x 30mm Screw, Part Number TBS030 The Black Widow Buttress Plates and Screws are designed for preventing migration or expulsion of allograft or autograft in the thoracolumbar to S1 Spinal region

Reason for recall

There is a problem with the mating feature between the plate and screw which could cause the plate opening not to interfere enough with the plate and head of the screw.

Distribution pattern

Nationwide Distribution to IN, LA, and TX

Code information

Part Number TBS020 Lot Number 0825R, NM01836, WT090314A, 09281, 09282, 09285, WT040412A, 08344, 09154, 09281R, NM01836R; Part Number TBS025 Lot Number 09161, 09286R, WT090314B, WT112613B, 08218, 08337, 09287R, NM01837, 08216; Part Number TBS030 Lot Number 08218, WT011711B, WT071014A

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0887-2017
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Amendia, Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0887-2017. Report a data issue: hello@regdatalab.com.