Z-0327-2018 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

Recall Details

Product Type
Devices
Report Date
January 17, 2018
Initiation Date
December 12, 2017
Termination Date
June 13, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product Description

EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.

Reason for Recall

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

Distribution Pattern

One medical device distributor in California.

Code Information

Lot 252564