Z-0327-2018 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

FDA device recall Z-0327-2018 was initiated by Encore Medical, Lp on Dec 12, 2017 and is designated Class II. Reason for recall: The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended. The recall status is terminated (terminated Jun 13, 2022). Affected quantity: 2 units.

Recall details

Product Type
Devices
Report Date
Jan 17, 2018
Initiation Date
Dec 12, 2017
Termination Date
Jun 13, 2022
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2 units

Product description

EMPOWR Porous Knee System TIBIAL PUNCH, SMALL, REF 801-05-202, NON STERILE, Qty 1 For use in orthopedic surgery.

Reason for recall

The Locking Punch Guide was not sufficiently guiding the large press fit punch, causing the implant to be placed 2-3mm more posterior than intended.

Distribution pattern

One medical device distributor in California.

Code information

Lot 252564