Z-0347-2027 Class II Ongoing

Recalled by SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION — Kyoto

FDA device recall Z-0347-2027 was initiated by SHIMADZU CORPORATION MEDICAL SYSTEMS DIVISION on Sep 30, 2025 and is designated Class II. Reason for recall: The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation. The recall status is ongoing. Affected quantity: 118 unts.

Recall details

Product Type
Devices
Report Date
Nov 11, 2026
Initiation Date
Sep 30, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
118 unts.

Product description

Trinias Opera Smart (C16s, F12s)/Opera Grand(C12s,C16s, F12s)/Smart OBL Interventional Fluoroscopic X-Ray System containing C-arm support (MH-500, MH-60),

Reason for recall

The mounting bolts securing the C-arm drive unit of the X-ray system may not be torqued to specification, potentially compromising structural stability and safe operation.

Distribution pattern

US Nationwide distribution in the states of SC, TX, FL, TN, CA, FL, NY.

Code information

Serial number/Shimadzu Medical Systems #/UDI: M1F0D3BF4001/ SMS20250167/ (01)04540217067145(11)250; M1F0CC4EB001/ SMS20250098/ (01)04540217067022(11)241; M1F0D27EA001/ SMS20250001/ (01)04540217067138(11)241; 41F0F43E7001/ SMS20250034/ (01)04540217068951(11)240; M1F0D3BE7001/ SMS20250248/ (01)04540217067145(11)240; M1F0D3BE6001/ SMS20240196/ (01)04540217067145(11)240; 41F0F40E5001/ N/A/ (01)04540217068937(11)240; 41F0F41E5001/ N/A/ (01)04540217068944(11)240; 41F0F43E5001/ SMS20240236/ (01)04540217068951(11)240; M1F0CD7E4001/ SMS20240162/ (01)04540217067046(11)240; 41F0F41E1001/ SMS20240062/ (01)04540217068944(11)240; 41F0F43DB002/ SMS20240101/ (01)04540217068951(11)231; 41F0F43DB001/ SMS20240100/ (01)04540217068951(11)231; M1F0CD7D4001/ SMS20230178/ (01)04540217067046(11)230

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0347-2027
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0347-2027. Report a data issue: hello@regdatalab.com.