Z-0380-2027 Class II Ongoing
Recalled by TransMedics, Inc.
— Andover, MA
FDA device recall Z-0380-2027 was initiated by TransMedics, Inc. on May 15, 2026 and is designated Class II. Reason for recall: The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying… The recall status is ongoing. Affected quantity: 6.
Recall details
- Product Type
- Devices
- Report Date
- Nov 18, 2026
- Initiation Date
- May 15, 2026
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 6
Product description
OCS Heart Perfusion Set Model/Catalog Number: 1200 Software Version: rev 3 Product Description: OCS Heart Perfusion Module, included in the Heart Perfusion Set used for organ transport.
Reason for recall
The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.
Distribution pattern
Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.
Code information
1. Perfusion Module SN: 4822-8-22505; Description: OCS Heart Perfusion Set ; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22505/14D20290409C*; 2. Perfusion Module SN: 4822-8-23601; Description: OCS Heart Perfusion Set; EXP Date: 7/27/29; UDI: *+B00948225/$$+74822-8-23601/14D20290727D*; 3. Perfusion Module SN: 4822-8-22496; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22496/14D20290409L*; 4. Perfusion Module SN: 4822-8-23497; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-23497/14D20290409N*; 5. Perfusion Module SN: 4822-8-22479; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22479/14D20290409M*; 6. Perfusion Module SN: 4822-8-22504; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22504/14D20290409B*; 7. Perfusion Module SN: 4822-8-23604; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23604/14D20290713B*; 8. Perfusion Module SN: 4822-8-23600; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23600/14D202907137*; 9. Perfusion Module SN: 4822-8-23499; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23499/14D20290713N*; 10. Perfusion Module SN: 4822-8-23602; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23602/14D202907139*; 11. Perfusion Module SN: 4822-8-23674; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23674/14D20290713I*; 12. Perfusion Module SN: 4822-8-22460; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22460/14D20290409C*; 13. Perfusion Module SN: 4822-8-23675; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23675/14D20290713J*; 14. Perfusion Module SN: 4822-8-22481; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22481/14D20290409F*; 15. Perfusion Module SN: 4822-9-22492; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-9-22492/14D20290409I*;
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0380-2027
- openFDA file date
- Nov 24, 2026
- Loaded by RegDataLab
- Nov 25, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for TransMedics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works