Z-0380-2027 Class II Ongoing

Recalled by TransMedics, Inc. — Andover, MA

FDA device recall Z-0380-2027 was initiated by TransMedics, Inc. on May 15, 2026 and is designated Class II. Reason for recall: The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying… The recall status is ongoing. Affected quantity: 6.

Recall details

Product Type
Devices
Report Date
Nov 18, 2026
Initiation Date
May 15, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
6

Product description

OCS Heart Perfusion Set Model/Catalog Number: 1200 Software Version: rev 3 Product Description: OCS Heart Perfusion Module, included in the Heart Perfusion Set used for organ transport.

Reason for recall

The Perfusion Module may have an unprogrammed Front End Board (FEB) which will not allow the system to proceed beyond the initial startup process. The OCS system will display an error message saying Unsupported Perfusion Module Present and will prevent further system use.

Distribution pattern

Worldwide - US Nationwide distribution in the states of CA, FL, IL, MA, MN, NC, NE, NY, TX and the country of Serbia;.

Code information

1. Perfusion Module SN: 4822-8-22505; Description: OCS Heart Perfusion Set ; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22505/14D20290409C*; 2. Perfusion Module SN: 4822-8-23601; Description: OCS Heart Perfusion Set; EXP Date: 7/27/29; UDI: *+B00948225/$$+74822-8-23601/14D20290727D*; 3. Perfusion Module SN: 4822-8-22496; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22496/14D20290409L*; 4. Perfusion Module SN: 4822-8-23497; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-23497/14D20290409N*; 5. Perfusion Module SN: 4822-8-22479; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22479/14D20290409M*; 6. Perfusion Module SN: 4822-8-22504; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22504/14D20290409B*; 7. Perfusion Module SN: 4822-8-23604; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23604/14D20290713B*; 8. Perfusion Module SN: 4822-8-23600; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23600/14D202907137*; 9. Perfusion Module SN: 4822-8-23499; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23499/14D20290713N*; 10. Perfusion Module SN: 4822-8-23602; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23602/14D202907139*; 11. Perfusion Module SN: 4822-8-23674; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23674/14D20290713I*; 12. Perfusion Module SN: 4822-8-22460; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22460/14D20290409C*; 13. Perfusion Module SN: 4822-8-23675; Description: OCS Heart Perfusion Set; EXP Date: 7/13/29; UDI: *+B00948225/$$+74822-8-23675/14D20290713J*; 14. Perfusion Module SN: 4822-8-22481; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-8-22481/14D20290409F*; 15. Perfusion Module SN: 4822-9-22492; Description: OCS Heart Perfusion Set; EXP Date: 4/9/29; UDI: *+B00948225/$$+74822-9-22492/14D20290409I*;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0380-2027
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for TransMedics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0380-2027. Report a data issue: hello@regdatalab.com.