Z-0389-2027 Class II Ongoing

Recalled by TransMedics, Inc. — Andover, MA

FDA device recall Z-0389-2027 was initiated by TransMedics, Inc. on Nov 20, 2025 and is designated Class II. Reason for recall: Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the… The recall status is ongoing. Affected quantity: 1179 (Liver Perfusion Sets).

Recall details

Product Type
Devices
Report Date
Nov 18, 2026
Initiation Date
Nov 20, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1179 (Liver Perfusion Sets)

Product description

OCS Liver System for organ transport OCS Liver Perfusion Set Model/Catalog Number: 3200 Product Description: The recalled product is a standard off the shelf syringe. It is used as part of solution delivery cassette (SDS Cassette) to be used with the perfusion module of OCS Liver System. OCS Liver Perfusion Sets within the serial number range between 4470-3-18076 and 4470-3-20343 may contain SDS cassettes within the impacted lot range of 4459-6-737116 to 4459-6-739965. Each Liver Perfusion Set contains three SDS cassettes. The BD syringe is a component of SDS Cassette PN 100004459, which is an accessory of the Liver Perfusion Set.

Reason for recall

Faulty Solution Delivery System (SDS) Cassettes within the device set were assembled with syringes that were manufactured with an atypical lubricant, which results in sub-par performance and has the potential to cause leaks.

Distribution pattern

Worldwide - US Nationwide distribution in the state of NE and the country of Italy.

Code information

OCS Liver Perfusion Set Serial Numbers from 4470-3-18076 to 4470-3-20343;

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0389-2027
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for TransMedics, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0389-2027. Report a data issue: hello@regdatalab.com.