Z-0382-2019 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North)

FDA device recall Z-0382-2019 was initiated by Randox Laboratories Ltd. on Oct 2, 2018 and is designated Class II. Reason for recall: Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no… The recall status is terminated (terminated May 30, 2019). Affected quantity: 2108.

Recall Details

Product Type
Devices
Report Date
Nov 14, 2018
Initiation Date
Oct 2, 2018
Termination Date
May 30, 2019
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
2108

Product Description

Liquid Cardiac Control, CQ5053

Reason for Recall

Randox has previously issued a recall for Liquid Cardiac Control CQ5053 lot 4245CK under REC334 on June 8, 2018. The firm has now confirmed the Liquid Cardiac Controls CQ5051, CQ5052 and CQ5053 are no longer suitable for the control of the Troponin T assay due to unacceptable variation between vials.

Distribution Pattern

Distributed to IN, NY, PR, and WV.