Z-0391-2018 Class II Terminated

Recalled by Philips Medical Systems (Cleveland) Inc — Cleveland, OH

Recall Details

Product Type
Devices
Report Date
January 24, 2018
Initiation Date
November 10, 2017
Termination Date
November 5, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3

Product Description

Ingenuity Core 128, (Code 728323) Intended to produce images of the head and body by computer reconstruction of x-ray transmission data taken at different angles and planes.

Reason for Recall

Philips has identified that certain fasteners cannot be confirmed to have been torqued according to specification. If the fasteners were not torqued as specified and a failure occurs while the system is in use, the parts secured by the fasteners may become loose, which may have a potential for harm.

Distribution Pattern

Worldwide Distribution -- US, Argentina, China, Republic of Korea, Russian Federation, Slovakia, Turkey, and United Kingdom. 4 in US, (1 military), 14 foreign accounts

Code Information

Code 728323, SN # 320394, UDI: (01)00884838059474(21)320394; 320395, UDI: (01)00884838059474(21)320395; and 336286.