Z-0398-2024 Class II Ongoing

Recalled by LEONI CIA CABLE SYSTEMS — Chartres, N/A

Recall Details

Product Type
Devices
Report Date
November 29, 2023
Initiation Date
October 6, 2023
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1 unit

Product Description

ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00003

Reason for Recall

Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.

Distribution Pattern

US Nationwide distribution in the states of MA, TN.

Code Information

UDI-DI: (01)03700383510017 SN: 100000001141