Z-0399-2024 Class II Ongoing

Recalled by LEONI CIA CABLE SYSTEMS — Chartres, N/A

FDA device recall Z-0399-2024 was initiated by LEONI CIA CABLE SYSTEMS on Oct 6, 2023 and is designated Class II. Reason for recall: Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an… The recall status is ongoing. Affected quantity: 3 units.

Recall details

Product Type
Devices
Report Date
Nov 29, 2023
Initiation Date
Oct 6, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3 units

Product description

ORION System with software OSS v6.0 and OSS v7.0- An electro-mechanical robotic arm for patient positioning in Radiation Therapy, Radiology, or Radiology image-guided treatment environment. Part Number: 1SOPT00009

Reason for recall

Users of the patient positioning system ORION System can be faced with a sudden short-term acceleration or with an unintended short-distance movement when the movements are again allowed after an unforeseen interruption signal of motion may result in a collision of the patient with another stationary component of the treatment room.

Distribution pattern

US Nationwide distribution in the states of MA, TN.

Code information

UDI-DI:(01)03700383510079 SN 100000004639 SN 100000004698 SN 100000004695

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0399-2024
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for LEONI CIA CABLE SYSTEMS? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0399-2024. Report a data issue: hello@regdatalab.com.