Z-0418-2020 Class II Terminated

Recalled by Cincinnati Sub-Zero Products LLC, A Gentherm Company — Cincinnati, OH

FDA device recall Z-0418-2020 was initiated by Cincinnati Sub-Zero Products LLC, A Gentherm Company on Jul 2, 2019 and is designated Class II. Reason for recall: Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the… The recall status is terminated (terminated Jun 22, 2020). Affected quantity: 3726.

Recall details

Product Type
Devices
Report Date
Nov 27, 2019
Initiation Date
Jul 2, 2019
Termination Date
Jun 22, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3726

Product description

Blanketrol II Hyper-Hypothermia System, Models 222R and 222S - Product Usage: The system is composed of a heater, a compressor, a circulating pump and blankets/pads.

Reason for recall

Warnings statements have been added- exceeding 40 degree C for extended periods may cause tissue damage. Additionally, a caution was added and clarifications were made regarding the use of the Automatic modes

Distribution pattern

Worldwide distribution - US Nationwide and countries of Afghanistan, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Belgium, Bhutan, Brazil, Brunei, Canada Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, France, Germany, Guatemala, Hong Kong, India, Indonesia, Israel, Italy, Japan, Kazakhstan, Kenya, Kuwait, Lebanon, Malaysia, Mexico, Morocco, Nepal, Netherlands, New Zealand, Oman, Pakistan, Palestine, Panama, Philippines, Qatar, Saudi Arabia, Singapore, South Africa, South Korea, Spain, Taiwan, Thailand, Trinidad and Tobago, Turkey, United Arab Emirates, Venezuela, Vietnam.

Code information

Model 222R (071-18490 through 123-20534) Model 222S (104-2-00001 through 161-2-01848)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0418-2020
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0418-2020. Report a data issue: hello@regdatalab.com.