Z-0825-2019 Class II Terminated
FDA device recall Z-0825-2019 was initiated by Cincinnati Sub-Zero Products LLC, A Gentherm Company on Dec 19, 2018 and is designated Class II. Reason for recall: Evidence supports that when customers performed a fuse replacement as part of a March 2017 field action, the fuse clips became stretched and were no longer mechanically tight with the fuse, creating… The recall status is terminated (terminated Apr 24, 2020). Affected quantity: 396.
Recall details
- Product Type
- Devices
- Report Date
- Feb 20, 2019
- Initiation Date
- Dec 19, 2018
- Termination Date
- Apr 24, 2020
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 396
Product description
Reason for recall
Distribution pattern
Worldwide Distribution - US Nationwide in the states of: AL, AZ, CA, CO, FL, GA, IL, IN, KY, LA, MA, MD, MI, MN, MO, NC, NJ, NY, OH, OK, PA, SC, TN, TX, VA, and WI. The products were distributed to the following foreign countries: Columbia and Taiwan.