Z-0479-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 2, 2020
- Initiation Date
- June 19, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 28 units
Product Description
EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.
Reason for Recall
There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.
Distribution Pattern
US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.
Code Information
all lot numbers