Z-0479-2021 Class II Ongoing

Recalled by Encore Medical, LP — Austin, TX

Recall Details

Product Type
Devices
Report Date
December 2, 2020
Initiation Date
June 19, 2020
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
28 units

Product Description

EMPOWR Partial Knee Peg Drill Guide; Model: 800-06-007.

Reason for Recall

There have been issues experienced with the peg preparation steps, specifically drilling of the Femoral posterior peg. Upon drilling, the shaft portion of the drill may bind within the Posterior Peg drill guide. The extended length of the drill guide and tight tolerances to ensure accurate preparation seem to be contributing to this situation. There is a potential of a delay in surgery if this issue were to recur.

Distribution Pattern

US: AZ, CA, FL, GA, IL, IN, MI, MN, NY, TX, VA, and WA.. No OUS Consignees.

Code Information

all lot numbers