Z-0495-2013 Class II Terminated

Recalled by Philips Healthcare Inc. — Andover, MA

FDA device recall Z-0495-2013 was initiated by Philips Healthcare Inc. on Nov 9, 2012 and is designated Class II. Reason for recall: Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms. The recall status is terminated. Affected quantity: 3165.

Recall details

Product Type
Devices
Report Date
Dec 19, 2012
Initiation Date
Nov 9, 2012
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
3165

Product description

Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.

Reason for recall

Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.

Distribution pattern

Worldwide Distribution - USA (nationwide) and Canada.

Code information

PIIC 865435 M3140 Information Center Low Acuity 865436 M3150 Information Center Local DB 865417 M3145 Information Center Small NW DB 865418 M3155 Information Center NW DB 865415 M3151 Information Center Client 865421  M3170 Patient Link 865416  M3177 Trend Display 865422  M3140 Upgrade 865423  M3150 Upgrade 865426  M3145 Upgrade 865427  M3155 Upgrade 865424  M3151 Upgrade 865266 - M3140 Information Center Low Acuity 865267  M3150 Information Center Local DB 865269  M3145 Information Center Small NW DB 865268  M3155 Information Center NW DB 865270  M3151 Information Center Client 865274  M3170 Patient Link 865271  M3177 Trend Display 865277  M3140 Upgrade 865278  M3150 Upgrade 865281  M3145 Upgrade 865282  M3155 Upgrade 865279  M3151 Upgrade PIIC iX 866023  IntelliVue Info Center iX 866117  PIIC Classic Upgrade 866027  PIIC iX HW Upgrade

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0495-2013
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Philips Healthcare Inc. →

Source: openFDA record for Z-0495-2013. Report a data issue: hello@regdatalab.com.