Z-0495-2013 Class II Terminated
Recalled by Philips Healthcare Inc.
— Andover, MA
FDA device recall Z-0495-2013 was initiated by Philips Healthcare Inc. on Nov 9, 2012 and is designated Class II. Reason for recall: Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms. The recall status is terminated. Affected quantity: 3165.
Recall details
- Product Type
- Devices
- Report Date
- Dec 19, 2012
- Initiation Date
- Nov 9, 2012
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3165
Product description
Philips IntelliVue Information Center/Client and Philips IntelliVue Information Center iX The intended use of the Information Center Software is to display physiologic waves, parameters, and trends, format data for strip chart recordings and printed reports, and provide the secondary annunciation of alarms from other networked medical devices at a centralized location. The Information Center Software provides for the retrospective review of alarms, physiologic waves and parameters from its database. An additional intended use of the Information Center Software is to provide primary annunciation of alarms and configuration and control access for networked telemetry monitors.
Reason for recall
Internal audio cable may have diminished retention force due to damaged contacts, which could lead to intermittent or loss of audible alarms.
Distribution pattern
Worldwide Distribution - USA (nationwide) and Canada.
Code information
PIIC 865435 M3140 Information Center Low Acuity 865436 M3150 Information Center Local DB 865417 M3145 Information Center Small NW DB 865418 M3155 Information Center NW DB 865415 M3151 Information Center Client 865421 M3170 Patient Link 865416 M3177 Trend Display 865422 M3140 Upgrade 865423 M3150 Upgrade 865426 M3145 Upgrade 865427 M3155 Upgrade 865424 M3151 Upgrade 865266 - M3140 Information Center Low Acuity 865267 M3150 Information Center Local DB 865269 M3145 Information Center Small NW DB 865268 M3155 Information Center NW DB 865270 M3151 Information Center Client 865274 M3170 Patient Link 865271 M3177 Trend Display 865277 M3140 Upgrade 865278 M3150 Upgrade 865281 M3145 Upgrade 865282 M3155 Upgrade 865279 M3151 Upgrade PIIC iX 866023 IntelliVue Info Center iX 866117 PIIC Classic Upgrade 866027 PIIC iX HW Upgrade
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0495-2013
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Philips Healthcare Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Philips Healthcare Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Philips Healthcare Inc. →
Use it in ChatGPT
Get a free API key
How the connector works