Z-0501-2022 Class II Terminated

Recalled by Draeger Medical, Inc. — Telford, PA

Recall Details

Product Type
Devices
Report Date
January 26, 2022
Initiation Date
December 16, 2021
Termination Date
April 17, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
5

Product Description

Draeger Fabius MRI Anesthesia Machine, Catalog Number 8607300

Reason for Recall

26 Fabius MRI Anesthesia machines were distributed with internal hoses routing Oxygen to the Oxygen Flush and Oxygen Flow Control interchanged. If the hospital O2 Central supply pressure is not stable, the O2 fresh gas flow could deviate from the adjusted value. If the deviation is not noticed, insufficient oxygenation of the patient may occur.

Distribution Pattern

US Nationwide distribution in the states of NY, PA, and WI.

Code Information

UDI 04048675041498 Serial Numbers ASPJ-0019 ASPJ-0020 ASPJ-0021 ASPK-0001 ASPK-0002