Z-0510-2024 Class II Ongoing

Recalled by Ortho-Clinical Diagnostics, INc. — Rochester, NY

FDA device recall Z-0510-2024 was initiated by Ortho-Clinical Diagnostics, INc. on Oct 25, 2023 and is designated Class II. Reason for recall: Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not… The recall status is ongoing. Affected quantity: 622 units 300 Slides (5 cartridge containing 60 slides).

Recall details

Product Type
Devices
Report Date
Dec 13, 2023
Initiation Date
Oct 25, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
622 units 300 Slides (5 cartridge containing 60 slides)

Product description

VITROS Chemistry Products ECO2 Slides-For in vitro diagnostic use only. VITROS Chemistry Products ECO2 Slides quantitatively measure total carbon dioxide (CO2) concentration in serum and plasma using VITROS 250/350/5,1 FS/4600 XT 3400 Chemistry Systems and the VITROS 5600/ XT 7600 Integrated System. Bicarbonate/carbon dioxide measurements are used in the diagnosis and treatment of numerous potentially serious disorders associated with changes in body acid-base balance. Product Code: 8262396

Reason for recall

Ortho Clinical Diagnostics (QuidelOrtho) confirmed during routine testing that some lots from Coating 3598 of VITROS ECO2 slides included carts which may contain consecutive slides that may not function as intended. These affected slides (potentially intermittent, consecutive slides within a cart or across consecutive carts) have an abnormality which may cause positively or negatively biased total carbon dioxide (CO2) results leading to either overdiagnosis of acidosis or missing alkalosis, both which may prompt unnecessary therapies such as IV fluids and/or bicarbonate or overdiagnosis of alkalosis which may also result in unnecessary use of IV fluids and electrolyte replacement or failure to detect acidosis and necessary treatment.

Distribution pattern

Nationwide Foreign: Bermuda Canada

Code information

UDI-DI: 10758750004645 Lot Numbers/Exp. Date: 5732-3598-4649 01-Feb-2025; 5732-3598-4650 01-Feb-2025; 5732-3598-4651 01-Feb-2025; 5732-3598-5102 01-Mar-2025; 5732-3598-5105 01-Mar-2025; 5732-3598-5189 01-Mar-2025

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0510-2024
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Ortho-Clinical Diagnostics, INc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Ortho-Clinical Diagnostics, INc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Ortho-Clinical Diagnostics, INc. →

Source: openFDA record for Z-0510-2024. Report a data issue: hello@regdatalab.com.