FDA device recall Z-0516-2016 was initiated by Ameditech Inc on Nov 16, 2015 and is designated Class III. Reason for recall: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the… The recall status is terminated (terminated Nov 9, 2016). Affected quantity: 3,411 kits.
Recall details
- Product Type
- Devices
- Report Date
- Dec 30, 2015
- Initiation Date
- Nov 16, 2015
- Termination Date
- Nov 9, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 3,411 kits
Product description
ProScreen 10 Panel Cup, Item No. PSCup-10M
Reason for recall
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Distribution pattern
Nationwide Distribution.
Code information
142596 142984 142985 143144 143145 143146 143240 143241 143351 143875 144102 144106 144239 144241 150651 151706 152165 152167 152168 153102 153107 153251 153411 153412 153633