Z-0518-2022 Class II Terminated

Recalled by GE Healthcare, LLC — Waukesha, WI

FDA device recall Z-0518-2022 was initiated by GE Healthcare, LLC on Nov 4, 2021 and is designated Class II. Reason for recall: The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer… The recall status is terminated (terminated Jul 24, 2026). Affected quantity: 12,526 total reported units (7 units of this model).

Recall details

Product Type
Devices
Report Date
Feb 2, 2022
Initiation Date
Nov 4, 2021
Termination Date
Jul 24, 2026
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
12,526 total reported units (7 units of this model)

Product description

GE Vscan Extend, being sold as "Vscan Extend Console"

Reason for recall

The firm identified that the Vscan Extend device will not power on because of depleted battery, due to the device not being turned off prior to battery removal or when being connected to a computer without AC power.

Distribution pattern

Domestic: AK, AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT. NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY, PR, and DC. International: Albania, Australia, Austria, Bahrain, Bangladesh, Belgium, Bosnia, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Dubai, Ecuador, El Salvador, Estonia, Fiji, Finland, France, Germany, Greece, Haiti, Hungary, Iceland, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Korea, Kuwait, Lebanon, Lithuania, Malaysia, Malta, Mexico, Moldova, Morocco, Myanmar, Netherlands, New Zealand, Norway, Papua New Guinea, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, Somalia, South Africa, Spain, Sweden, Switzerland, Thailand, Togo, Trinidad and Tobago, Turkey, Ukraine, United Arab Emirates, United Kingdom, Vietnam, and Western Sahara.

Code information

Information reads "Serial Number (UDI, Model Number)". VH1622261Y (01008406821216511118042621VH1622261Y, 5693455), VH161664XA (01008406821216511118031921VH161664XA, 5693455), VH2627194S (01008406821216511120071321VH2627194S, 5693455), VH160129E8 (01008406821216511117021721VH160129E8, 5693455), VH160834YQ (01008406821216511117110021VH160834YQ, 5693455), VH1642124F (01008406821216511119040921VH1642124F, 5693455), VH150152W7 (01008406821216511116080021VH150152W7, 5693455), VH160113JQ (01008406821216511117020321VH160113JQ, 5693455), VH161979EY (01008406821216511118030021VH161979EY, 5693455), VH162446PC (01008406821216511118061221VH162446PC, 5693455), VH161984GS (01008406821216511118030021VH161984GS, 5693455), VH161970GQ (01008406821216511118030021VH161970G, 5693455), VH1619899S (01008406821216511118030021VH1619899S, 5693455), VH160237FQ (01008406821216511117030021VH160237FQ, 5693455), VH1600005U (5693455), VH160007CX (5693455), VH160009UF (5693455), VH150153FK (5693455), VH161672SF (5693455), VH1624281N (5693455), VH150103AI (5693455), VH150150Z6 (5693455), VH160786JM (5693455), VH1608086P (5693455), VH1600107T (5693455), VH1607384V (5693455), VH1600112G (5693455), VH160012JB (5693455), VH1607406X (5693455), VH160742EX (5693455), VH1608215J (5693455), VH160633NC (5693455), VH160025HU (5693455), VH160744ZM (5693455), VH150151QA (5693455), VH160730NB (5693455), VH161629J2 (5693455), VH161631UU (5693455), VH160735W7 (5693455), VH161598QQ (5693455), VH1600189D (5693455), VH1501544T (5693455), VH160017E2 (5693455), VH150123AH (5693455), VH163240NI (5693455), VH160005HB (5693455), VH1656702Z (5693455), VH16567191 (5693455), VH1656729W (5693455), VH1656736Y (5693455), VH165674RL (5693455), VH165675AQ (5693455), VH165676D7 (5693455), VH1656777R (5693455), VH165678YC (5693455), VH165679H1 (5693455), VH150112VD (5693455), VH150113IF (5693455), VH1607343V (5693455), VH160012JB (5693455), VH150101S2 (5693455), VH160001S1 (5693455), VH160003N9 (5693455), VH1616079K (5693455), VH150105RI (5693455), VH150118BJ (5693455), VH1501144B (5693455), VH150119KV (5693455), VH150122PV (5693455), VH150104JE (5693455), VH150067PS (5693455), VH150102JE (5693455), VH16162756 (5693455), VH260391G7 (5693455), VH2609418F (5693455), VH26096618 (5693455), VH1600301G (5693455), VH160021FJ (5693455), VH160163HW (5693455)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0518-2022
openFDA file date
Nov 27, 2026
Loaded by RegDataLab
Nov 28, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for GE Healthcare, LLC? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for GE Healthcare, LLC: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about GE Healthcare, LLC →

Source: openFDA record for Z-0518-2022. Report a data issue: hello@regdatalab.com.