Z-0519-2016 Class III Terminated
FDA device recall Z-0519-2016 was initiated by Ameditech Inc on Nov 16, 2015 and is designated Class III. Reason for recall: Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the… The recall status is terminated (terminated Nov 9, 2016). Affected quantity: 1,989 kits.
Recall details
- Product Type
- Devices
- Report Date
- Dec 30, 2015
- Initiation Date
- Nov 16, 2015
- Termination Date
- Nov 9, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,989 kits
Product description
ProScreen 10 Panel Dip Card w/Adult, Item No. PSDA-10MO, PSDA-10MMO300, PSDA-10PPX, PSDA-10BUP, PSD-12BUP, PSD-10M, PSCupA-6MB-W, PSCup-10MMO300, PSD-5AB, PSD-10CMO3, PSCupA-10MO, PSCupA-12M-W I-DXA-1127-023, PSCupA-12TBU.
Reason for recall
Ameditechs Drugs of Abuse Tests listed above have shown reduced reactivity for up to two (2) of the nineteen (19) Benzodiazepine (BZO) compounds for which specificity information is provided in the product insert. The particular BZO compounds for which there may be reduced reactivity are Clonazepam and Chlordiazepoxide.
Distribution pattern
Nationwide Distribution.
Code information
144902 150474 151686 151687 152043 152225 152312 152402 152952 153306 153563 153663 153664 153665 153680 153682 153703 153704 153759 153760 153763 153852 153853