Z-0543-2024 Class II Ongoing
Recalled by Ortho-Clinical Diagnostics, Inc.
— Rochester, NY
FDA device recall Z-0543-2024 was initiated by Ortho-Clinical Diagnostics, Inc. on Oct 10, 2023 and is designated Class II. Reason for recall: During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred… The recall status is ongoing. Affected quantity: 63 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 20, 2023
- Initiation Date
- Oct 10, 2023
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 63 units
Product description
VITROS 3600 Immunodiagnostic System Product Code 6802783 and Refurbished Product Code 6802914 Running Software Version 3.8.0
Reason for recall
During planned monitoring of data post-release of Software Version 3.8.0 an increased occurrence of software issues relating to unresponsive subsystems, unresponsive user interface (sometimes referred to as a screen freeze), and/or printer disconnections was identified. The consequence of these issues would be a delay in reporting results due to the required system shutdowns to recover the system. This delay in reporting results could occur during any assay, including which the speed of the result is deemed critical or STAT and could result in a delay of treatment in a patient. Due to the release strategy of this software version, the user could not update the system but a field engineer had to manually install each affected system.
Distribution pattern
US Nationwide. Global Distribution.
Code information
Product Code: 6802783, UDI-DI: 10758750002979; Product Code: 6802914, UDI-DI: 10758750007103; Serial Numbers: J36000109 J36000294 J36001012 J36001489 J36000292 J36000340 J36001024 J36001495 J36000484 J36000348 J36001027 J36001513 J36000489 J36000446 J36001051 J36001515 J36000955 J36000511 J36001088 J36001519 J36001151 J36000517 J36001108 J36001528 J36001167 J36000580 J36001111 J36001535 J36001222 J36000688 J36001112 J36001570 J36001283 J36000753 J36001117 J36001573 J36001330 J36000861 J36001237 J36001575 J36001348 J36000877 J36001245 J36001579 J36000920 J36001246 J36001581 J36000923 J36001401 J36001582 J36000929 J36001403 J36001583 J36000933 J36001427 J36001600 J36000941 J36001451 J36001602 J36000942 J36001488 J36001618 J36001000
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0543-2024
- openFDA file date
- Nov 27, 2026
- Loaded by RegDataLab
- Nov 28, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Ortho-Clinical Diagnostics, Inc.? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.
Question to ask in Claude
Use RegDataLab to give me an FDA risk summary for Ortho-Clinical Diagnostics, Inc.: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.
Ask Claude about Ortho-Clinical Diagnostics, Inc. →
Use it in ChatGPT
Get a free API key
How the connector works