Z-0559-2019 Class II Terminated

Recalled by Tosoh Bioscience Inc — Grove City, OH

FDA device recall Z-0559-2019 was initiated by Tosoh Bioscience Inc on Aug 23, 2018 and is designated Class II. Reason for recall: The firm has implemented new product labeling changes for the Tosoh AIA-360 Analyzer. These labeling changes affect the Tosoh Quick Reference Guide, Training Manual and Training DVD previously… The recall status is terminated (terminated Oct 20, 2020). Affected quantity: 1228.

Recall details

Product Type
Devices
Report Date
Dec 12, 2018
Initiation Date
Aug 23, 2018
Termination Date
Oct 20, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1228

Product description

AIA-360 Analyzer, product code 019945/019945R The Tosoh AIA-360 Automated Immunoassay Analyzer is for use in vitro diagnostic to provide testing for a broad menu of assays in a unique patented dry reagent format.

Reason for recall

The firm has implemented new product labeling changes for the Tosoh AIA-360 Analyzer. These labeling changes affect the Tosoh Quick Reference Guide, Training Manual and Training DVD previously provided by Tosoh Bioscience. These changes are being made to align the Tosoh Quick Reference Guide and associated training materials to the Operators Manual and provide clarification to the requirement that the maintenance of the B/F probe is to be performed by Tosoh a field service representative, the requirement for use of 70% Ethanol during daily shutdown procedures and the requirement for use of CAP Class 1 reagent grade water for dilution of the concentrated wash and diluent solutions and reconstitution of lyophilized reagents. Failure to follow these requirements may negatively affect system performance, lead to corrosion of instrument components over time, or result in bacterial contamination, respectively.

Distribution pattern

Worldwide Distribution: US (Nationwide) and countries of: Mexico, Canada, Colombia, Chile, Venezuela, Ecuador, Uruguay, Peru, and Honduras.

Code information

Change to labeling affects all devices.

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0559-2019
openFDA file date
Nov 19, 2026
Loaded by RegDataLab
Nov 20, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Tosoh Bioscience Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0559-2019. Report a data issue: hello@regdatalab.com.