Z-0562-2018 Class II Terminated
Recalled by Philips Medical Systems (Cleveland) Inc
— Cleveland, OH
FDA device recall Z-0562-2018 was initiated by Philips Medical Systems (Cleveland) Inc on Dec 11, 2017 and is designated Class II. Reason for recall: A software issue causes PET reconstructions to fail intermittently. It was determined that reconstructions fail due to a negative table position (-1 value is inserted) in the raw data list file… The recall status is terminated (terminated Aug 30, 2022). Affected quantity: 24.
Recall details
- Product Type
- Devices
- Report Date
- Feb 21, 2018
- Initiation Date
- Dec 11, 2017
- Termination Date
- Aug 30, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 24
Product description
Ingenuity TF PET/CT (model 882442) running software version 4.0.2 This device is a diagnostic imaging system that combines Positron Emission Tomography (PET) and X-ray Computed Tomography (CT) systems. The CT subsystem produces cross-sectional images of the body by computer reconstruction of x-ray transmission data. The PET subsystem produces images of the distribution of PET radiopharmaceuticals in the patient body (specific radiopharmaceuticals are used for whole body, brain, heart and other organ imaging). CT data is applied to the PET data for attenuation correction. The PET subsystem also provides for list mode, dynamic, and gated acquisitions. This system is intended for patients of all ages
Reason for recall
A software issue causes PET reconstructions to fail intermittently. It was determined that reconstructions fail due to a negative table position (-1 value is inserted) in the raw data list file, rather than the actual table position. This error has been found to occur in two scenarios: 1) When the system operator cancels an acquisition a. The error will occur every time a scan is cancelled by the operator. 2) Couch position requests within the software sequence were delayed a. The error occurs intermittently, but has been found to occur more frequently when the gantry s network is heavily loaded with multiple retrospective reconstructions running in parallel. b. The error manifests to the technologist by an error message and Failed status on the Reconstruction Monitor and the error message Result {0} failed to reconstruct" on the Acquisition Workflow window during reconstruction after the patient scan has been completed. In both scenarios, the acquisition data will not be able to be reconstructed and will therefore be unusable.
Distribution pattern
Worldwide Distribution - US Distribution to the states of : CA, OH, IL, PA and NY.
Code information
Ingenuity TF PET/CT (model 882442) SN: 2003 2004 2005 2006 2007 2008 2009 2010 2011 2012 2013 2015 2016 2017 2018 2019 2020 2021 2022 2023 2024 2025 2026 2027
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0562-2018
- openFDA file date
- Nov 25, 2026
- Loaded by RegDataLab
- Nov 26, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Philips Medical Systems (Cleveland) Inc? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works