Z-0604-2016 Class II Terminated

Recalled by Elekta, Inc. — Atlanta, GA

FDA device recall Z-0604-2016 was initiated by Elekta, Inc. on Dec 18, 2015 and is designated Class II. Reason for recall: When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and… The recall status is terminated (terminated Sep 15, 2021). Affected quantity: 154.

Recall details

Product Type
Devices
Report Date
Jan 13, 2016
Initiation Date
Dec 18, 2015
Termination Date
Sep 15, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
154

Product description

Oncentra External Beam - VMAT - Radiation therapy planning system Product Usage: The Oncentra system is a radiation treatment planning software designed to analyze and plan radiation treatments in three dimensions for the purpose of treating patients with cancer.

Reason for recall

When using the option "Tumor Overlap Fraction" in VMAT planning it has been observed that in rare cases the system does include an organ at risk as target volume. This could result in open MLC, and open jaws in areas away from the target volume.

Distribution pattern

Worldwide Distribution - US Nationwide in the states of CA, NY, PA, and the countries of Austria, Belarus, Canada, China, Denmark, Finland, France, Germany, Greece, India, Italy, Japan, Netherlands, Norway, Poland, South Korea, Spain, Sweden and United Kingdom.

Code information

Versions 3.3 and higher

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0604-2016
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0604-2016. Report a data issue: hello@regdatalab.com.