Z-0606-2025 Class II Ongoing
FDA device recall Z-0606-2025 was initiated by GE Healthcare (China) Co., Ltd. on Nov 5, 2024 and is designated Class II. Reason for recall: GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly… The recall status is ongoing. Affected quantity: 4 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 11, 2024
- Initiation Date
- Nov 5, 2024
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 4 units
Product description
GE Healthcare Revolution Ascend Elite, Model Number 6967000-100; X-ray/computed tomography system
Reason for recall
GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.
Distribution pattern
Worldwide distribution.
Code information
UDI/DI 00195278718129, Serial/Sales Order Numbers: CBFLG2400002HM, CBFLG2400005HM, CBFLG2400004HM, CBFLG2400003HM