Z-1241-2025 Class II Ongoing

Recalled by GE Healthcare (China) Co., Ltd. — Beijing, N/A

FDA device recall Z-1241-2025 was initiated by GE Healthcare (China) Co., Ltd. on Jan 31, 2025 and is designated Class II. Reason for recall: For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly… The recall status is ongoing. Affected quantity: 33.

Recall details

Product Type
Devices
Report Date
Mar 5, 2025
Initiation Date
Jan 31, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
33

Product description

Optima MR360, NMRI system

Reason for recall

For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.

Distribution pattern

Domestic US distribution nationwide. International distribution worldwide.

Code information

A. UDI-DI 00840682113762 System ID 0910274038 B. No UDI-DI System IDs 082427040134 082427040185 082427060043 082427070088 082427070093 082427070099 082427100174 082427100178 082427100197 082427120140 082427140095 082427220046 082427270041 082427310088 082427310092 083027288790310 0856270061 AE1373MR01 DZ1339MR03 FI1432MR01 H4678MR04 MRR9282 MRR9877 PL2059MR01 PL2433MR01 RU3338MR03 SA1056MR03 SA2191MR04 SA2253MR01 SA2352MR01 TN1008MR01 TN1241MR01 34352MRS01 (Update 12/29/25 - Additional System IDs 83027809699615, LV4026MR01) (Update 1/28/26 - Additional System ID 82427190104)

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1241-2025
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1241-2025. Report a data issue: hello@regdatalab.com.