Z-0605-2025 Class II Ongoing

Recalled by GE Healthcare (China) Co., Ltd. — Beijing

Recall Details

Product Type
Devices
Report Date
December 11, 2024
Initiation Date
November 5, 2024
Termination Date
N/A
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
16 units

Product Description

GE Healthcare Revolution Ascend, Model number 6969000-100; X-ray/computed tomography system

Reason for Recall

GE HealthCare has become aware of a potential issue on certain Discovery, Optima, Revolution series CT systems, and certain operator console upgrades that can result in rotated CT images. Particularly with helical, cine, and cardiac (if applicable) scan modes, after the first image of an exam, each subsequent image can be rotated by an increasing amount. The amount of progressive image rotation is dependent on the gantry speed and the duration of the scan. The final image in the exam can be rotated up to a maximum of ~56 degrees. The entire anatomy rotates as a whole by equal degrees, without distortion, and the anatomical positioning and z-axis of the images remains correct. Small degrees of image rotation might not be noticed. If image rotation is noticed and necessitates a rescan of the patient, the same image rotation will reoccur.

Distribution Pattern

Worldwide distribution.

Code Information

UDI/DI 00840682146173, Serial/Sales Order Numbers: CBDWG2400109HM, CBDWG2400103HM, CBDWG2400102HM, CBDWG2400096HM, CBDWG2400111HM, CBDWG2400106HM, CBDWG2400115HM, CBDWG2400113HM, CBDWG2400105HM, CBDWG2400112HM, CBDWG2400116HM, CBDWG2400114HM, CBDWG2400107HM, CBDWG2400104HM, CBDWG2400110HM, CBDWG2400108HM