Z-0608-2013 Class III Terminated
FDA device recall Z-0608-2013 was initiated by Radiometer America Inc on Jun 6, 2006 and is designated Class III. Reason for recall: The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45. The recall status is terminated (terminated Feb 5, 2013). Affected quantity: 840 Ampules.
Recall details
- Product Type
- Devices
- Report Date
- Jan 2, 2013
- Initiation Date
- Jun 6, 2006
- Termination Date
- Feb 5, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 840 Ampules
Product description
Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.
Reason for recall
The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Distribution pattern
Distributed to KY, NC, and NY.
Code information
Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.