Z-0608-2013 Class III Terminated

Recalled by Radiometer America Inc — Westlake, OH

FDA device recall Z-0608-2013 was initiated by Radiometer America Inc on Jun 6, 2006 and is designated Class III. Reason for recall: The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45. The recall status is terminated (terminated Feb 5, 2013). Affected quantity: 840 Ampules.

Recall details

Product Type
Devices
Report Date
Jan 2, 2013
Initiation Date
Jun 6, 2006
Termination Date
Feb 5, 2013
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
840 Ampules

Product description

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

Reason for recall

The barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.

Distribution pattern

Distributed to KY, NC, and NY.

Code information

Lot R0040, expiration date February 2008; Lot R0042, expiration date April 2008.