Z-0628-2016 Class III Terminated
FDA device recall Z-0628-2016 was initiated by Becton, Dickinson And Company, BD Biosciences on Dec 14, 2015 and is designated Class III. Reason for recall: Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100%… The recall status is terminated (terminated Oct 31, 2016). Affected quantity: 1,388 units.
Recall details
- Product Type
- Devices
- Report Date
- Jan 20, 2016
- Initiation Date
- Dec 14, 2015
- Termination Date
- Oct 31, 2016
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1,388 units
Product description
FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.
Reason for recall
Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.
Distribution pattern
Nationwide Distribution
Code information
Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0628-2016
- openFDA file date
- Nov 16, 2026
- Loaded by RegDataLab