Z-0628-2016 Class III Terminated

Recalled by Becton, Dickinson And Company, BD Biosciences — San Jose, CA

FDA device recall Z-0628-2016 was initiated by Becton, Dickinson And Company, BD Biosciences on Dec 14, 2015 and is designated Class III. Reason for recall: Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100%… The recall status is terminated (terminated Oct 31, 2016). Affected quantity: 1,388 units.

Recall details

Product Type
Devices
Report Date
Jan 20, 2016
Initiation Date
Dec 14, 2015
Termination Date
Oct 31, 2016
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1,388 units

Product description

FMC7 FITC, Catalog No. 340918 Analyte Specific Reagent.

Reason for recall

Becton, Dickinson and Company (BD) has determined that the FMC7 FITC (ASR) is exhibiting functional degradation. The functional degradation has the effect of reducing the percent positive from 100% down to approximately 87%.

Distribution pattern

Nationwide Distribution

Code information

Lot No. 4162520, 4206855, 4267919, 4276818, 4309977, 5009530, 5036624, 5085573, 5182818

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0628-2016
openFDA file date
Nov 16, 2026
Loaded by RegDataLab
Nov 17, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0628-2016. Report a data issue: hello@regdatalab.com.