Z-0660-2019 Class II Terminated

Recalled by Randox Laboratories Ltd. — Crumlin (North)

Recall Details

Product Type
Devices
Report Date
January 2, 2019
Initiation Date
November 19, 2018
Termination Date
April 23, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
250 kits (32 kits US / 218 kits OUS)

Product Description

SP CAL (LIQ), REF IT2692 This product is intended for in vitro diagnostic use with Randox Alpha-1-Antitrypsin, Alpha-1-Acid Glycoprotein, IgA, IgG and IgM assays that require sample predilution.

Reason for Recall

Calibrator value sheets have been modified for assigned value of Alpha-1-Acid-Glycoprotein (AGP), IgA and IgG to improve the alignment. IgG measurement reference has also been corrected.

Distribution Pattern

US Distribution to states of : IN, NJ, NY, TX and WV and internationally to Ireland.

Code Information

05055273204049 Lot Number: 415879 Date of Expiry - 28 March 2019