Z-0717-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 24, 2024
- Initiation Date
- December 4, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 236 units
Product Description
Soltive SuperPulsed Laser System TFL FIBER CLEAVER, Model Number TFL-AFC
Reason for Recall
The strippers and cleavers are provided non-sterile and labelled as reusable/autoclavable. Olympus does not have validated cleaning and sterilization instructions. Use of a non-sterile fiber stripper or cleaver on a sterile fiber poses a risk of contamination.
Distribution Pattern
US Nationwide. Global Distribution
Code Information
Lot Number: All units sold prior to 2023