Z-0723-2026 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- December 3, 2025
- Initiation Date
- October 17, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 19
Product Description
IntelePACS - InteleConnect / TechPortal
Reason for Recall
Software application that receives digital images and data to be communicated, processed, manipulated, enhanced, stored, displayed has a bug that could cause data loss in Relevant Clinical Info field when modifying studies through Case Editor in either Technologist Portal or Referring Physician Portal, which could result in loss of clinical information, which could impact clinical decision-making.
Distribution Pattern
US Nationwide distribution in the states of AZ, CT, AL, IL, WA, MD, NJ,CA, TN, IN, FL, NC.
Code Information
UDI-DI: B228INTELEPACS0. Version/Dates of Distribution: PACS-5-6-1-R64 through R68/02 Jul 2025 - 10 Sep 2025, PACS-5-7-1-R42 through R47/26 June 2025 - 25 Sep 2025, PACS 5-8-1-R20 through R23/26 Jun 2025 - 25 Sep 2025, PACS-5-9-1-R1 through R3/30 Jun 2025 - 09 Sep 2025