Z-1208-2025 Class II Ongoing
FDA device recall Z-1208-2025 was initiated by INTELERAD MEDICAL SYSTEMS INCORPORATED on Feb 5, 2025 and is designated Class II. Reason for recall: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT… The recall status is ongoing. Affected quantity: 88.
Recall details
- Product Type
- Devices
- Report Date
- Mar 5, 2025
- Initiation Date
- Feb 5, 2025
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 88
Product description
IntelePACS (Image Fusion Module) - InteleViewer
Reason for recall
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Distribution pattern
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Code information
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later