Z-1208-2025 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- March 5, 2025
- Initiation Date
- February 5, 2025
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 88
Product Description
IntelePACS (Image Fusion Module) - InteleViewer
Reason for Recall
A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.
Distribution Pattern
Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia
Code Information
UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later