Z-1208-2025 Class II Ongoing

Recalled by INTELERAD MEDICAL SYSTEMS INCORPORATED — Montreal, N/A

FDA device recall Z-1208-2025 was initiated by INTELERAD MEDICAL SYSTEMS INCORPORATED on Feb 5, 2025 and is designated Class II. Reason for recall: A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT… The recall status is ongoing. Affected quantity: 88.

Recall details

Product Type
Devices
Report Date
Mar 5, 2025
Initiation Date
Feb 5, 2025
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
88

Product description

IntelePACS (Image Fusion Module) - InteleViewer

Reason for recall

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

Distribution pattern

Worldwide - US Nationwide distribution in the states of AZ, NC, MI, FL, NH, WY, CA, SC, AL, AR, IN, MD, CO, TN, IL, NJ, OH, MN, WV, KE, NE, VA, WA, GE, WI and the countries of New Zealand, Canada, Australia

Code information

UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-1208-2025
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-1208-2025. Report a data issue: hello@regdatalab.com.