Z-0726-2024 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 24, 2024
- Initiation Date
- August 29, 2023
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 179
Product Description
PRIMA TT Glenoid Monoblock/Modular reverse TT baseplates: Description/Product Code: BASEPLATE D.25MM REGULAR/1975.14.500, BASEPLATE D.28MM REGULAR/1975.14.800, BASEPLATE D.25MM FULL WEDGED 10-degrees/1975.14.510, BASEPLATE D.28MM FULL WEDGED 10-degrees/1975.14.810, BASEPLATE D.25MM FULL WEDGED 15-degrees/1975.14.615, BASEPLATE D.28MM FULL WEDGED 15-degrees/1975.14.815, BASEPLATE D.28MM FULL WEDGED 15-degrees X/1975.14.865, BASEPLATE D.25 MM FULL WEDGED 20-degrees/1975.14.620, BASEPLATE D.28 MM FULL WEDGED 20-degrees/1975.14.820, BASEPLATE D.28 MM FULL WEDGED 20-degrees X/1975.14.870
Reason for Recall
Potential for baseplate peripheral holes to be out of specification due to a manufacturing issue, which could lead to loosening of the implant, or the screw moving from its original position.
Distribution Pattern
US: IL, MO, CA, IN, MI, OK
Code Information
Product Code/UDI-DI: 1975.14.500/08033390244824, 1975.14.800/08033390244893, 1975.14.510/08033390231299, 1975.14.810/08033390244909, 1975.14.615/08033390244848, 1975.14.815/08033390244916, 1975.14.865/08033390244930, 1975.14.620/08033390231312, 1975.14.820/08033390244923, 1975.14.870/08033390244947