Z-0744-2025 Class II Ongoing

Recalled by Limacorporate S.P.A — San Daniele Del Friuli

FDA device recall Z-0744-2025 was initiated by Limacorporate S.P.A on Oct 2, 2024 and is designated Class II. Reason for recall: Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery. The recall status is ongoing. Affected quantity: ~87 devices (U.S.).

Recall details

Product Type
Devices
Report Date
Jan 1, 2025
Initiation Date
Oct 2, 2024
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
~87 devices (U.S.)

Product description

TEMA Elbow - Axle Small REF 1590.15.010 TEMA Elbow - Axle Large REF 1590.15.020 TEMA Elbow System is a cemented total elbow prosthesis.

Reason for recall

Due to a potential difference in mechanical behavior between the implanted device and tool(s) used during revision surgery.

Distribution pattern

U.S: FL and TX O.U.S.: N/A

Code information

Product/Catalog Numbers: 1590.15.010 TEMA ELBOW - AXLE SMALL UDI-DI code: 08033390128865 Lot/Serial Numbers: 23031752300054 23177372300230 23173852400087 Product/Catalog Number: 1590.15.020 TEMA ELBOW - AXLE LARGE UDI-DI code: 08033390128872 Lot/Serial Numbers: 22276962300054 23325482400047 24006192400193

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0744-2025
openFDA file date
Nov 24, 2026
Loaded by RegDataLab
Nov 25, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Limacorporate S.P.A? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.

Get a free API key

How the connector works

Source: openFDA record for Z-0744-2025. Report a data issue: hello@regdatalab.com.