Z-0729-2017 Class II Terminated

Recalled by Merge Healthcare, Inc. — Hartland, WI

FDA device recall Z-0729-2017 was initiated by Merge Healthcare, Inc. on Mar 28, 2016 and is designated Class II. Reason for recall: If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the… The recall status is terminated (terminated Mar 20, 2020). Affected quantity: 198 sites potentially have the effected versions.

Recall details

Product Type
Devices
Report Date
Dec 21, 2016
Initiation Date
Mar 28, 2016
Termination Date
Mar 20, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
198 sites potentially have the effected versions

Product description

Merge Cardio software. Merge Cardio is a system intended to be used to acquire, store, print, transfer, and archive clinical information including images, Hemodynamic studies and reports, measurements (via import from DICOM Structured Reporting, text files or optical character recognition of measurements captured on images) and cardiology signal (waveform) data.

Reason for recall

If images are sent without an order in the system, they will be matched with the latest order on the current patient/modality matching potentially resulting in the matching of the report to the incorrect accession number (but still associated to the correct patient).

Distribution pattern

Nationwide Distribution

Code information

Versions: 8.30.0, 8.30.1, 9.0, 9.0.1, 9.0.2, 9.0.3, 9.0.4, 9.0.5, 9.0.6, 9.0.7, 9.0.8, 9.0.9, 10.0, 10.0.1, 10.1, and 10.1.1

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0729-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0729-2017. Report a data issue: hello@regdatalab.com.