Z-0739-2018 Class I Terminated

Recalled by Ventana Medical Systems Inc — Oro Valley, AZ

Recall Details

Product Type
Devices
Report Date
March 7, 2018
Initiation Date
December 19, 2017
Termination Date
November 22, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
358 units

Product Description

ultraView SISH DNP Detection Kit, Ventana Part Number 800-098, Roche GMMI 05907136001 in vitro diagnostic

Reason for Recall

Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.

Distribution Pattern

Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,

Code Information

UDI 04015630972135, Lot Number: Y17990