Z-0746-2013 Class III Terminated
FDA device recall Z-0746-2013 was initiated by Alere San Diego, Inc. on Oct 22, 2012 and is designated Class III. Reason for recall: Out of range (increased) HDL Cholesterol results. The recall status is terminated (terminated Sep 26, 2013). Affected quantity: 11,166 kits.
Recall details
- Product Type
- Devices
- Report Date
- Feb 6, 2013
- Initiation Date
- Oct 22, 2012
- Termination Date
- Sep 26, 2013
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 11,166 kits
Product description
Alere Cholestech LDX Multianalyte Control, 2 vial set, Catalog Number: 88769 Alere Cholestech LDX Multianalyte Control (hereafter referred to as Controls) enables users to monitor the performance of total cholesterol (TC), high density lipoprotein cholesterol (HDL), triglycerides (TRG), glucose (GLU), alanine aminotransferase (ALT), and aspartate aminotransferase (AST) test procedures on the Alere Cholestech LDX System. The results obtained for the controls are to be compared with the assigned values given on the Expected Values Card, accompanying the package insert, to determine if the procedure is within control limits. Assayed quality control material for use with the Alere Cholestech LDX System.
Reason for recall
Out of range (increased) HDL Cholesterol results.
Distribution pattern
Worldwide Distribution - USA (nationwide) Puerto Rico and South Korea.
Code information
Lot#C2510, C2510A, C2572, C2572A, C2593, and C2593A.