Z-0748-2023 Class II Ongoing
Recalled by Heartware, Inc.
— Miami Lakes, FL
FDA device recall Z-0748-2023 was initiated by Heartware, Inc. on Nov 29, 2022 and is designated Class II. Reason for recall: Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may… The recall status is ongoing. Affected quantity: 73 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 28, 2022
- Initiation Date
- Nov 29, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 73 units
Product description
HeartWare HVAD Pump Kit, REF 1104JP
Reason for recall
Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may be delayed, which may result in patient harm from a prolonged pump stop, if an urgent driveline connection or controller issue needs to be addressed. In the event a hardened driveline cover is observed, contact your Medtronic field representative for further assessment and/or servicing.
Distribution pattern
Worldwide distribution.
Code information
Serial Numbers: HW40657, HW40658, HW40659, HW40660, HW40662, HW40663, HW40664, HW40665, HW40666, HW40667, HW40668, HW40789, HW40816, HW40881, HW40950, HW40951, HW40953, HW40954, HW40956, HW40957, HW40958, HW40960, HW40961, HW41045, HW41047, HW41048, HW41050, HW41214, HW41251, HW41252, HW41254, HW41255, HW41256, HW41258, HW41259, HW41260, HW41262, HW41263, HW41264, HW41265, HW41266, HW41269, HW41270, HW41271, HW41308, HW41309, HW41314, HW41316, HW41317, HW41670, HW41672, HW41673, HW41674, HW41675, HW41699, HW41700, HW41701, HW41783, HW41791, HW41792, HW41803, HW41804, HW41824, HW41828, HW41847, HW41848, HW42285, HW42722, HW42771, HW25169, HW25170, HW25172, HW25174
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0748-2023
- openFDA file date
- Nov 19, 2026
- Loaded by RegDataLab
- Nov 20, 2026
These dates say when openFDA published the data file and when RegDataLab last loaded it.
Need the full FDA record for Heartware, Inc.? Ask Claude or ChatGPT with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. The API key is free: no card.
Get a free API key
How the connector works