Z-0748-2024 Class II Ongoing

Recalled by Olympus Corporation Of The Americas — Center Valley, PA

FDA device recall Z-0748-2024 was initiated by Olympus Corporation Of The Americas on Dec 1, 2023 and is designated Class II. Reason for recall: Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential… The recall status is ongoing. Affected quantity: 82,200 total units (5 per box).

Recall details

Product Type
Devices
Report Date
Jan 24, 2024
Initiation Date
Dec 1, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
82,200 total units (5 per box)

Product description

SOLTIVE SuperPulsed Laser Fibers Model Numbers TFL-FBX150S, TFL-FBX200S, TFL-FBX365S, TFL-FBX550S, TFL-FBX940S, TFL-FBX150BS, TFL-FBX200BS

Reason for recall

Olympus identified inconsistencies in the Instructions for Use regarding cleaning the fiber tip and fiber cleaving and stripping instructions. Following the current instructions presents the potential for the fiber to break during use, which may cause burns, procedure delays, prolonged surgery, and foreign body in patient.

Distribution pattern

Worldwide distribution - US Nationwide and the countries of Canada, Europe, Hong Kong, Australia, Japan, Singapore, and Latin America.

Code information

Models UDI/GTIN TFL-FBX150S 00821925043916 TFL-FBX200S 00821925043930 TFL-FBX365S 00821925043992 TFL-FBX550S 00821925044036 TFL-FBX940S 00821925044074 TFL-FBX150BS 00821925043879 TFL-FBX200BS 00821925043978 All lots affected

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0748-2024
openFDA file date
Nov 25, 2026
Loaded by RegDataLab
Nov 26, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Need the full FDA record for Olympus Corporation Of The Americas? Ask Claude with the RegDataLab connector and get every site's inspections, Form 483s, warning letters, recalls and import alerts, each with a link to the FDA source. Free to start, no card.

Question to ask in Claude

Use RegDataLab to give me an FDA risk summary for Olympus Corporation Of The Americas: recent inspections and their outcomes, warning letters, recalls and import alerts by site, with source links and dates.

Ask Claude about Olympus Corporation Of The Americas →

Source: openFDA record for Z-0748-2024. Report a data issue: hello@regdatalab.com.