Z-0749-2023 Class II Ongoing
FDA device recall Z-0749-2023 was initiated by Heartware, Inc. on Nov 29, 2022 and is designated Class II. Reason for recall: Over time, the driveline cover may become hardened, creating difficulty or an inability to slide the cover back to access the driveline to controller connector. Access to the driveline connector may… The recall status is ongoing. Affected quantity: 126 units.
Recall details
- Product Type
- Devices
- Report Date
- Dec 28, 2022
- Initiation Date
- Nov 29, 2022
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 126 units
Product description
Reason for recall
Distribution pattern
Worldwide distribution.