Z-0758-2023 Class II Ongoing
Recalled by GE Medical Systems, LLC
— Waukesha, WI
Recall Details
- Product Type
- Devices
- Report Date
- December 28, 2022
- Initiation Date
- November 10, 2022
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 113 units
Product Description
Vivid i N ultrasound. Not marketed in the US. Used in ultrasound imaging and analysis in Fetal/Obstetrics; Abdominal/GYN; Pediatric; Small Organ (breast, testes, thyroid); Neonatal Cephalic; Adult Cephalic; Cardiac (adult and pediatric); Peripheral Vascular; Musculo-skeletal Conventional and Superficial; Urology (including prostate); Transesophageal; Transrectal; Transvaginal; and Intraoperative (abdominal, thoracic, and vascular).
Reason for Recall
Potential for batteries that have not been replaced at 2 years, as recommended in the Service Manual, they can fail or possibly emit smoke, or catch fire.
Distribution Pattern
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV and the countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, AZERBAIJAN, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, CHINA Colombia, CONGO, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, CZECH REPUBLIC, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Ethiopia, Finland, France, Georgia, Germany, Ghana, Greece, Guatemala, Honduras, HONG KONG, Hungary, ICELAND, INDIA, Indonesia, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Korea (Republic of), Kosovo, Kuwait, KYRGYZSTAN, Latvia, Lebanon, Libyan Arab Jamahiriya, Lithuania, LUXEMBOURG, Macedonia, MALAYSIA, Mali, Mauritania, Mauritius, Mexico, MICRONESIA, Moldova, Morocco, Mozambique, Namibia, Nepal, Netherlands, NEW ZEALAND, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, SYRIA, TAIWAN, TANZANIA, THAILAND, Trinidad and Tobago, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen.
Code Information
No US Distribution. Model Number H45561AC, System ID: DZ1237US02, AO1015US01, EG1843US01, EG5882US01, EG1621US04, EG1621US03, 2538978U01, 083038042857318, 083038602007811, 083038602826113, 0835384095, IQ1165US05, IQ1376US01, KE1049US01, KW1010US04, KW1102US04, KW1001US35, KW1011US21, KW1011US22, KW1006US14, KW1098US06, KW1020US15, KT1061US01, KT1062US02, LB1359US04, 1553UL1026, 0834380243, 0834384085, 0834384095, MA1251US13, MA1272US10, OM1042US03, OM1007US11, OM1007US12, PK1331US08, PK1331US06, PA1097US01, SA1201US02, SA1014US07, SA1337US05, SA1060US08, SA1147US05, SA1149US03, SA1051US09, SA2220US02, SA2220US01, SA2072US17, SA1028US06, SA1089US07, SA1138US05, SA2158US07, SA1340US02, SA1223US15, SA1002US08, SA1036US06, SA1025US03, SA2214US03, SA1272US10, SA1024US04, SA1101US04, SA1034US20, SA1249US11, SA1034US23, SA1058US60, SA1058US47, SA1058US59, SA1332US03, SA1030US05, SA1282US05, ZA2573US01, ZA2422US01, ZA2328US01, ZA2339US01, ZA2420US01, ZA1024US07, ZA2348US01, ZA2358US01, ZA1030US10, ZA2421US01, ZA1025US16, ZA1025US25, ZA2378US01, ZA2510US01, TN1028US04, 01002ULT25, 06004ULT19, 07021ULT02, P34189UL17, 21002ULT18, 23002ULT06, P23003UL12, 32004ULT18, 34013ULT13, 34037ULT16, 34013ULT14, P35076UL03, 38006ULT08, 42002ULT12, 42002ULT10, P50005UL02, 55001ULT13, 60001ULT07, P61029UL05, AE1283US01, AE1353US01, AE1016US16, AE1016US14, AE1016US15, AE1011US18, AE1331US01, AE1015US16, YE1071US04, YE1071US03