Z-0766-2017 Class II Terminated

Recalled by Encore Medical, Lp — Austin, TX

FDA device recall Z-0766-2017 was initiated by Encore Medical, Lp on Sep 19, 2016 and is designated Class II. Reason for recall: It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery. The recall status is terminated (terminated Oct 20, 2017). Affected quantity: 162 units.

Recall details

Product Type
Devices
Report Date
Dec 21, 2016
Initiation Date
Sep 19, 2016
Termination Date
Oct 20, 2017
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
162 units

Product description

Empowr PS Knee System Box Cut Guide The EMPOWER PS Box Cut Guide is used to make the housing resection for the post of the posterior stabilized implant

Reason for recall

It has been discovered that during Empowr PS Total Knee replacement, while chiseling for bone removal in the box area, femoral chipping has occurred in surgery.

Distribution pattern

Nationwide Distribution

Code information

202351L01, 202351L02, 202351L03, 202351L04, 202351L05, 202351L06, 202351L07, 202351L08, 202351L09, 203779L09, 202351L10, 203779L10, 203203L01

Source

Dataset
openFDA device recall enforcement reports
FDA record
This recall in openFDA (JSON)
Enforcement Report
Search FDA's Enforcement Report for Z-0766-2017
openFDA file date
Nov 10, 2026
Loaded by RegDataLab
Nov 11, 2026

These dates say when openFDA published the data file and when RegDataLab last loaded it.

Source: openFDA record for Z-0766-2017. Report a data issue: hello@regdatalab.com.