Z-0766-2020 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 8, 2020
- Initiation Date
- November 4, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 1
Product Description
Apollo EZ, Model Number 9784152036 - Product Usage: The Apollo EZ remote-controlled table is X-ray equipment that can be used to perform general radiographic and fluoroscopic examinations. The system is a fixed position type. The movements of the table and tube support are motorized and remote-controlled. Medical purpose is the diagnosis, visualization and documentation of various human diseases or injuries.
Reason for Recall
An amendment to the IFU and Service Manual is being issued to prescribe the replacement of the angulation chain after 8 years of service of the of the unit. After testing, it was found that due to metal fatigue under sever workload conditions, the angulation chain might fail before the expected lifetime of 10 years.
Distribution Pattern
US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, and Puerto Rico.
Code Information
Serial Numbers: 13101261