Z-1428-2017 Class II Terminated

Recalled by Villa Sistemi Medicali S.P.A. — Buccinasco, N/A

Recall Details

Product Type
Devices
Report Date
March 15, 2017
Initiation Date
December 27, 2016
Termination Date
March 9, 2021
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
65 devices in total

Product Description

Apollo; Models: 9784000131 9784000231 9784020131 9784020231 9784100231 9784120131 9784120231 9784130131 Product Usage: General, multi-function x-ray systems capable of routine radiographs and general fluoroscopy procedures.

Reason for Recall

Patient fell from the table because the footrest of the device detached from the table top while the patient was positioned on the table and table was moving from vertical position.

Distribution Pattern

US Nationwide distribution in the states of AL, AZ, CA, CO, CT, DC, FL, IA, IL, IN, LA, MA, MD, MI, MN, MO, MS, NJ, NY, PA, Puerto Rico, TN, TX, UT, and W.I

Code Information

Serial Numbers: 06100127 06110138 07010149 07020158 07020163 07020165 07030171 07040206 09010566 09090605 10100752 11100873 15041446