Z-0780-2014 Class II Terminated
FDA device recall Z-0780-2014 was initiated by Elekta, Inc. on Nov 25, 2013 and is designated Class II. Reason for recall: When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the… The recall status is terminated (terminated May 13, 2014). Affected quantity: 13.
Recall details
- Product Type
- Devices
- Report Date
- Jan 29, 2014
- Initiation Date
- Nov 25, 2013
- Termination Date
- May 13, 2014
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 13
Product description
XiO RTP System The XiO RTP System is used to create treatment plans for any cancer patient for whom external beam radiation therapy or brachytherapy has been prescribed. The system will calculate and display, both on-screen and in hard-copy, either two or three-dimensional radiation dose distributions within a patient for a given treatment plan set-up.
Reason for recall
When the number of fractions is changed for a proton spot beam, the MU are recomputed and rounded if necessary, but the dose is not recalculated as it should be even after saving and re-opening the plan.
Distribution pattern
Worldwide Distribution - US Distribution including the states of IL, IN, LA, NJ, TX, WA and WI., and the countries of Belgium, Germany, Italy and Netherlands.
Code information
Model or Catalog No. MUJ; Software Version XiO Versions 4.80 and higher
Source
- Dataset
- openFDA device recall enforcement reports
- FDA record
- This recall in openFDA (JSON)
- Enforcement Report
- Search FDA's Enforcement Report for Z-0780-2014
- openFDA file date
- Nov 10, 2026