Z-0799-2014 Class II Terminated

Recalled by Zeiss, Carl Inc — Thornwood, NY

Recall Details

Product Type
Devices
Report Date
January 29, 2014
Initiation Date
December 20, 2013
Termination Date
April 8, 2015
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
335 units Total (14 domestically & 321 internationally)

Product Description

Compensator Slider Lambda 6x20 component (part number 473704-0000-000) utilized on all upright ZEISS microscopes (Axio Imager, Axio Scope. A1, Axio Lab. A1). Component used to determine composition of urinary calculi, urinary sediments, bio-crystallates and diagnosis of pathological characteristics in medical applications (e.g.: testing for gout or pseudogout); also used for non-medical scientific research.

Reason for Recall

Carl Zeiss Microscopy, LLC has recalled the Compensator Slider Lamda 6x20 component utilized on Standard Pathology Microscopes due to misalignment of the lambda plate.

Distribution Pattern

Worldwide Distribution - Nationwide including NY, TX, NC, FL, ME, MA and MI; and Internationally to Arab Emirates, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, China, Czech Republic, Egypt, France, Germany, Hong Kong, India, Italy, Kosovo, Lithuania, Mexico, The Netherlands, New Zealand, Peru, Poland, Spain, and Taiwan.

Code Information

part number 473704-0000-000