Z-2441-2018 Class II Terminated
Recall Details
- Product Type
- Devices
- Report Date
- July 25, 2018
- Initiation Date
- May 29, 2018
- Termination Date
- July 26, 2024
- Voluntary/Mandated
- FDA Mandated
- Product Quantity
- 8
Product Description
Disinfection unit UV (432332-9020-000) for the Celldiscoverer 7 microscope
Reason for Recall
Under certain circumstances, the firmware makes it possible for the Disinfection unit UV (432332-9020-000) to activate outside of the Celldiscoverer 7 housing. This may result in exposure of the users to harmful UV radiation.
Distribution Pattern
US Nationwide Distribution
Code Information
432332-9020-000