Z-0800-2021 Class II Terminated

Recalled by Ortho-Clinical Diagnostics, Inc. — Rochester, NY

Recall Details

Product Type
Devices
Report Date
January 13, 2021
Initiation Date
November 20, 2020
Termination Date
October 24, 2023
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
40 units

Product Description

enGen Laboratory Automation System, enGen Track System, Catalog Number ENGEN - Product Usage: intended for use in the in vitro quantitative measurement of a variety of analytes of clinical interest, using both VITROS Chemistry Products Slides (colorimetric endpoint, rate, ion-selective electrode, and immunorate methods) and VITROS Chemistry Products MicroTip liquid reagents (spectrophotometric and spectrophotometric immunoassay methods).

Reason for Recall

A software anomaly may cause an aliquoted sample to be labelled as the incorrect sample. This may lead to results being associated with the wrong patient sample and potentially lead to inappropriate intervention with the potential for injury to the patient.

Distribution Pattern

Worldwide distribution - US Nationwide distribution including the states of CA, CO, MA, MD, NC, OH, TX, VA, VT, and Belgium, Bermuda, Canada, China, France, Italy, Japan, Mexico, Norway, Portugal, Singapore, Spain, Sweden, and United Kingdom.

Code Information

enGen Laboratory Automation Systems and VITROS Automation Solutions Using All TCAutomationTM configured with Thermo Scientific Aliquoter & Labeller Module and using Software Versions 4.6 and below