Z-0812-2019 Class I Ongoing
FDA device recall Z-0812-2019 was initiated by Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) on January 17, 2019 and is designated Class I. Reason for recall: A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode wit… The recall status is ongoing. Affected quantity: 47086 units.
Recall Details
- Product Type
- Devices
- Report Date
- February 20, 2019
- Initiation Date
- January 17, 2019
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 47086 units
Product Description
vitatron (implantable pulse generator): (a) A-series, Model Numbers: A30A1, A60A1 (b) E-series, Model Numbers: E50A1, E60A1 (c) G-series, Model Numbers: G70A1, G70A2 (d) Q-series, Model Numbers: Q50A2, Q70A2, Q80A2
Reason for Recall
A subset of Medtronic dual chamber pacemakers distributed worldwide between 10 March 2017 and 7 January 2019 under the brand names Adapta, Versa, and Sensia when programmed to a dual chamber mode with atrial-sensing, may experience a circuit error that affects device functionality
Distribution Pattern
Wordlwide
Code Information
vitatron (implantable pulse generator): (a) A-series, Model Number: A30A1, GTIN 00643169709331; A60A1, GTIN 00643169709294, All Serial Numbers (b) E-series, Model Number: E50A1, GTIN 00643169709126, 00643169709140; E60A1, GTIN 00643169709164, 00643169709188, All Serial Numbers (c) G-series, Model Numbers: G70A1, GTIN 00643169709249, 00643169709263; G70A2, GTIN 00643169883000, All Serial Numbers (d) Q-series, Model Number: Q50A2, GTIN 00643169883017; Q70A2, GTIN 00643169882980; Q80A2, GTIN 00643169882997, All Serial Numbers