Z-0816-2021 Class II Ongoing
Recall Details
- Product Type
- Devices
- Report Date
- January 20, 2021
- Initiation Date
- November 7, 2020
- Termination Date
- N/A
- Voluntary/Mandated
- Voluntary: Firm initiated
- Product Quantity
- 8272 cases
Product Description
NaturaLyte Liquid Acid concentrate bottles, Model Number 08-2301-3 - Product Usage: used in conjunction with Fresenius Renal Technologies Sodium Bicarbonate Concentrate in a compatible 45X dilution three-stream artificial kidney (hemodialysis) machine.
Reason for Recall
The sodium and/or calcium levels in these concentrates were below the manufacturing specifications.
Distribution Pattern
US Nationwide distribution.
Code Information
Lot Numbers: 20LXAC056 20LXAC057 20NXAC038 20LXAC041 20LXAC055 Additional affected lots as of 3/17/21: 20NXAC013 20NXAC035