Z-0822-2020 Class II Terminated

Recalled by Isopure Corp — Louisville, KY

Recall Details

Product Type
Devices
Report Date
January 29, 2020
Initiation Date
April 19, 2019
Termination Date
May 29, 2020
Voluntary/Mandated
Voluntary: Firm initiated
Product Quantity
1

Product Description

Isopure Sodium Bicarbonate Mixing and Distribution System (IsoMix PRO) with a Finish Thompson Centrifugal Pump Model #110192-3

Reason for Recall

There is a potential for an uneven plastic encapsulation of the magnet ring during the molding process and partial axial displacement. This issue could cause uneven wear during use, cause exposure of the magnet to solution, and thereby result in corrosion and the presence of iron.

Distribution Pattern

The products were distributed to the following US states: AL, FL, GA, MA, MD, MN, NC, NM, NY, OH, OK, TX, and VA.

Code Information

IsoMix Pro UDI: (01)00852456007085; ISOMIX Serial Number IMP-838703 with Pump SN 180210D18